The XRROID Effect

 

This document is adapted from the QXCI system with permission from Professor William C. Nelson.

The word Xrroid is defined as the testing of a person's Electro Physiological Reactivity to thousands of substances at biological speeds. Biological speeds are defined as those approaching the ionic exchange speed of a persons' electrical reaction to the items in their immediate environment. This is a speed of approximately 1/100 of a second.

The Xrroid is the process of measuring the client reaction to such items as vitamins, homeopathics, enzymes, hormones, allersodes, isodes, nosodes and so on.

The Xrroid is the invention of Dr. Nelson and was first used in 1985 in the EPFX device of the Eclosion. This was registered with the FDA of America in 1989. The process has been greatly advanced technologically in the QXCI device. The Xrroid has been used on millions of clients around the world for over a decade.

The process has been clinically tested with results being published in medical journals and articles being presented in several world wide medical conferences. The users of the systems have sent in thousands of testimonials and reports of dramatic success come in daily. The device, when used as instructed, may mean seeing a client once a week.

 

 

The tickle of testing a person to thousands of items at fast speeds seems to promote an increase in the wellness of the EPR field that promotes a change or destabilization in the EPR field of the client. This will lead to inaccurate Xrroid test results for a period of up to 48 hours. So for this time the therapies can be done successfully but the Xrroid test will be less accurate.

There have been occasions during the last decade when the device has been used incorrectly by repeatedly checking tests on the same day. This appears to be the work of overly suspicious minds. They will report back to us with dismay for even though the first results are always accurate the second or third results seem not to be. Often these reports come from those who cling to older technology or have ulterior motives and often the reports have not been checked. However, recently when the Chinese distributor had a similar comment, the Chinese representative had an observation. Could it be that the Xrroid test might produce some effect on the EPR of the client?

Education Service in The UK and the world

 

 Clients will have hyper-reactivity states after testing. Some clients report heightened senses of taste, smell, coordination, flexibility, and even ESP. Some are not aware of the changes and other family members report noticing the differences. During this period the Xrroid retesting will often be inaccurate, but therapies can be used during this time. The recovery time appears to vary depending on the client condition and can be from 24 hours minimum to 100 hours maximum.

Our tests have shown that the Xrroid itself has healing effects since clients have demonstrated improved trivector patterns. Athletes consistently report heightened reflexes, improved coordination, and faster motor skills. One Xrroid test will result in several improvements such as clarity of thought process or eye hand coordination. However, after two or more Xrroid tests a state of hyperactivity can ensue for hours or days.

Please keep the Xrroid tests to a minimum. This change in EPR shows just how effective the Xrroid is.

 Education Service in The UK and the world


Reactivity Vs Resonance

This document is adapted from the QXCI system with permission from Professor William C. Nelson.

This article is designed to make the process of the QXCI /SCIO device more clear.

The measurement of Electro Physiological Reactivity (EPR) is clearly documented in energetic medicine. The majority of measuring devices designed to do this use skin impedance or resistance. These devices mostly all use point probes that depend on application by the therapist to the skin. The application of the point probe is subject to idiomotor or subtle muscle control. This allows the intervention of therapist control, which may effect the result. The same idiomotor control that makes the ouigi board appear to work, can effect the control of these energetic medical devices. If the point probe is applied slowly the reading will be lower than if the probe is applied quickly. It is the speed of delivery of the probe that controls the result more so than the pressure.

Is all of the use of this field of energetic medicine complete bunk? Even though there is a measure of operator interference, these devices have been found to measure something.

The body's reaction, or EPR, is made up of changes of voltage, amperage and resistance. Therefore the perfect measurement of EPR is found within changes in the trivector of amperage, voltage and resistance. However, measuring resistance alone would have some purpose.

The next difficulty with the antiquated point probe devices has been the limitation of the time of the test. The body's reaction to items is triggered by an ionic exchange. This is the basis of reaction since electron transport or ionic exchange dictates it. The speed of ionic exchange in the body is approximately 1/100 of a second. The point probes are not able to measure at this speed whereas the computerised QXCI device is designed to measure the total reaction at biological speeds. What are the old style point probe systems measuring? They are measuring the left over resonance of the initial reaction.

After the body has had it's initial reaction it will continue to react to the item for a variety of reasons in a variety of patterns. To show how a person could react, diagrams may be viewed within the simple help manual for the QXCI.

When we perform the Xrroid test we are measuring only the initial reaction and thus we can catalogue the priority of the initial reaction. To measure the residual resonance would take over one second per item and thus the Xrroid test would take over six hours. On completion of the test we can use the individual test button to measure any of the single items in the matrix. This takes about 2 seconds and will prompt a yellow panel displaying the initial reaction or Reactivity, (the Allergic reactivity), and the residual Resonance. The Resonance determines the extent to which a person needs the item in question. The point probe systems measure only the residual resonance whereas the QXCI device measures both with no operator interference.

 

 

I hope that you will find this information helpful in using the QXCI system.

The Coherence measurement is the percentage of how close the signal returns to the signal that is sent. This increases if there is a harmonic reaction in the client and any number above 75 indicates a positive reaction.

The Fourier number is the mathematical relationship number of the resulting 7 part harmonic dissection. This is a complicated mathematical concept which as yet does not seem to have any meaning. If you find one call us!